Modal | DP6150(A/B) | |||
Bandwidth(-3dB) | DP6150 | 70MHz | ||
DP6150A | 100MHz | |||
DP6150B | 200MHz | |||
Rise time | DP6150 | ≤5ns | ||
DP6150A | ≤3.5ns | |||
DP5150B | ≤1.75ns | |||
Accuracy | ±2% | |||
Range selection(Attenuation rate) | 50X/500X | |||
Maximum differential voltage(DC + Peak AC) | 50X | ±150V | ||
500X | ±1500V | |||
Common mode voltage(DC + Peak AC) | ±1500V | |||
Maximum differential mode VS frequency curve | Figure 1 | |||
Maximum input voltage-to-earth (Vrms) |
600V CATIII 1000V CATII | |||
Input impedance | Single-ended to ground |
5MΩ | ||
Between inputs |
10MΩ | |||
Input capacitance | Single-ended to ground | <4pF | ||
Between inputs | <2pF | |||
CMRR | DC | >80dB | ||
100kHz | >60dB | |||
1MHz | >50dB | |||
Noise(Vrms) | 50X | <50mV | ||
500X | <300mV | |||
Differential overvoltage detection level | 50X | ≥150V | ||
500X | ≥1500V | |||
Propagation time | Probe | About 9ns | ||
BNC Line(1m) | About 5ns | |||
Bandwidth limit filters (5MHZ) |
≥-3dB@5MHz | |||
Overload indicator (red light) | Yes | |||
Overload alarm | Yes(Can shut up manually) | |||
Automatic save | Yes | |||
Offset setting function | Yes( Set in test mode) | |||
Terminate load |
≥100kΩ | |||
Power supply | USB 5V/1A adapter | |||
Safety standard | EN61010-1: 2010 | |||
EMC standard | EN61326-1:2013 EN61000-3-2:2006+A1:2009+A2:2009 EN61000-3-3:2013 |